Valvular & structural heart diseaseMeta-analysis

1-Year Outcomes in a Pooled Cohort of Harmony Transcatheter Pulmonary Valve Clinical Trial Participants

JACC Cardiovasc IntervPublished

  • Pooled analysis of prospective clinical trials
  • 19 patients with a discontinued early device iteration evaluated separately

Authors' conclusion

The Harmony TPV device demonstrated favorable clinical and hemodynamic outcomes across studies and valve types through 1 year.

Study context

Population
Patients with severe pulmonary regurgitation (PR) (echocardiography) or PR fraction ≥ 30% (cardiac MRI) and clinical indications for pulmonary valve replacement, native or surgically repaired right ventricular outflow tract (RVOT)primary analysis
Participants
87 patients
Comparison
TPV22 vs TPV25 Harmony transcatheter pulmonary valve
  • TPV22n = 42
  • TPV25n = 45

Pooled result

As reported in the abstract

Freedom from PR, stenosis and reintervention · 1 year

Further pooled results

OutcomeTPV22TPV25
None/trace or mild PR · 1 yearNot reported in the abstract97%

Death · 1 year

No deaths

Safety

Nonsustained VT

Nonsustained VT, all patients
16%

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

1-Year Outcomes in a Pooled Cohort of Harmony Transcatheter Pulmonary Valve Clinical Trial Participants. JACC Cardiovasc Interv. Published 17 May 2023. doi:10.1016/j.jcin.2023.03.002