Valvular & structural heart disease

1-Year Real-World Outcomes of Transcatheter Aortic Valve Replacement (TAVR) With the Fifth-Generation Balloon-Expandable Valve in the United States

JACC Cardiovasc IntervPublished Retrospective cohort study

Authors' conclusion

TAVR with the S3UR is associated with superior 1-year clinical outcomes and lower gradients than its predecessors, with less PVL and low valve reintervention.

Study context

Population
Patients undergoing native TAVR, STS/ACC TVT Registry, September 2022 to March 2023matched from 32,536 S3 and S3U
Participants
4,598 vs 4,598
Exposure
TAVR with S3UR vs S3 and S3U valves
  • S3URn = 4,598
  • S3 and S3Un = 4,598reference
Outcome
Clinical and echocardiographic outcomes
Follow-up
1 year
Methodological detailspropensity-score matched, national registry (STS/ACC TVT Registry), Median follow-up 358 days (Q1-Q3: 47-365) S3UR vs 364 days (Q1-Q3: 80-365) S3 and S3U
  • Propensity-score matched
  • National registry (STS/ACC TVT Registry)
  • Median follow-up 358 days (Q1-Q3: 47-365) S3UR vs 364 days (Q1-Q3: 80-365) S3 and S3U

Main finding

As reported in the abstract

Clinical and echocardiographic outcomes

S3UR

All-cause mortality
7.6%
Mild or greater paravalvular leak (PVL)
15.6%

S3 and S3U

All-cause mortality
9.7%
Mild or greater PVL
18.5%

All-cause mortality

S3UR7.6%
S3 and S3U9.7%

Mild or greater paravalvular leak (PVL)

S3UR15.6%
S3 and S3U18.5%
S3UR vs S3 and S3UHazard ratio (95% CI)P value
All-cause mortality
0.8(0.67–0.93)
P = 0.004
Mild or greater PVL
0.82(0.69–0.97)
P = 0.02
Larger effective orifice areas with S3UR at discharge
P < 0.0001
Lower mean gradient with S3UR at discharge, maintained at 30 days and 1 year
P < 0.0001

Observational data: associations, not effects.

Further findings

Predictors of 1-year mortality

S3UR and mild or greater PVL predictive of 1-year mortality in the overall cohort and in low-risk patients

Safety

OutcomeS3URS3 and S3UHR (95% CI)P value
Life-threatening bleeding · 1 year · propensity-matched2.0%2.7%0.7 (0.54–0.94)P = 0.03
Valve reintervention · 1 year · propensity-matched0.6%0.4%1.46 (0.77–2.78)P = 0.25

Valve reintervention · 1 year

Valve reintervention remained rare at 1 year

Original publication

1-Year Real-World Outcomes of TAVR With the Fifth-Generation Balloon-Expandable Valve in the United States. JACC Cardiovasc Interv. Published 24 November 2024. doi:10.1016/j.jcin.2024.11.015

Summarised from the published abstract; every number is checked against it

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.