Valvular & structural heart diseaseRandomised controlled trial

3-Year Clinical Outcomes Comparing the Amulet vs Watchman FLX for Left Atrial Appendage Closure in Patients With Atrial Fibrillation

J Am Coll CardiolPublished

  • Randomised controlled trial
  • investigator-initiated
  • multicentre (8 centres)
  • 3 patients randomised to Amulet received Watchman FLX
  • 3-year follow-up rate 96.4% (Amulet) and 97.3% (Watchman)

Authors' conclusion

At 3 years after LAAC, there was no significant difference in the ischemic risk between the Amulet and the Watchman FLX groups.

Study context

Population
Patients with AF and high bleeding risk undergoing left atrial appendage closure (LAAC)220 completed LAAC
Participants
221 patients
Comparison
Amulet vs Watchman FLX for percutaneous LAAC
Primary endpoint
Composite of:
  • CV death
  • Stroke
  • TIA
  • Systemic embolism
Follow-up
3 years

Primary result

Composite ischemic endpoint

Hazard ratio

0.58

95% CI 0.33–1.03

P = 0.06

By group

Amulet vs Watchman
HR 0.58 (95% CI 0.33–1.03); P = 0.06

As reported in the abstract

Further findings

OutcomeAmuletWatchmanHR (95% CI)P value
Composite ischemic endpoint, as-treated population · 3 years · as-treated17.0%31.1%0.53 (0.30–0.96)P = 0.035
Composite ischemic endpoint, per-protocol population · 3 years · per-protocol16.2%29.2%0.54 (0.29–1.00)P = 0.049

Original publication

3-Year Clinical Outcomes Comparing the Amulet vs Watchman FLX for Left Atrial Appendage Closure in Patients With Atrial Fibrillation. J Am Coll Cardiol. Published 8 July 2025. doi:10.1016/j.jacc.2025.03.535

Summarised from the published abstract; every number is checked against it