Arrhythmias & electrophysiology

Accuracy of Implantable Loop Recorders: Multicenter, Multidevice Comparison

JACC Clin ElectrophysiolPublished multicenter, multivendor database

Authors' conclusion

ILRs demonstrate variable accuracy in arrhythmia detection, with a substantial burden of false-positive alerts across all vendors despite artificial intelligence-based and other algorithmic enhancements. Further improvement of alert algorithms to reduce clinic burdens is needed.

Study context

Population
Patients with implantable loop recorders (medtronic (MDT) LINQ II, BSX LUX-Dx, Biotronik (BIO) Biomonitor III, Abbott (ABT) Confirm Rx), January 2022 to March 2023random sample (∼283 per device) from a database of 437,351 electrocardiogram-verified episodes from 6,756 patients
Participants
1,140 patients
Exposure
Arrhythmia detection accuracy among currently used implantable loop recorders (ILRs)
Outcome
Total alerts
Methodological detailsrandom sample selected based on a prospectively determined power analysis, alerts independently adjudicated by trained physicians, 25 centers
  • Random sample selected based on a prospectively determined power analysis
  • Alerts independently adjudicated by trained physicians
  • 25 centers

Main finding

As reported in the abstract

Total alerts

25,826

Alerts per patient

median of 5 alerts

IQR: 2–14

Observational data: associations, not effects.

Further findings

Atrial tachycardia/AF alerts (share of alerts)

PPV

OutcomeMDTBSXBIOABT
PPV after adjudication0.560.730.230.51
Bradycardia detection PPVMDT primarily due to undersensing0.360.96— not stated in the abstract— not stated in the abstract
Pause detection PPV— not stated in the abstract0.70— not stated in the abstract0.01
Tachycardia alert PPV0.290.610.620.27

— not stated in the abstract

  • Pause detection PPV: ABT 79.8% due to undersensing

Causes of false-positive atrial tachycardia/atrial fibrillation (AF) alerts with BIO devices

Ectopy
75.8%
Combined ectopy with oversensing/noise
5.3%
Isolated oversensing
5.2%
Sinus arrhythmias
3.9%
Indeterminate reason
3.7%
Noise
3.6%
Bradycardia
2.5%

Original publication

Accuracy of Implantable Loop Recorders: Multicenter, Multidevice Comparison. JACC Clin Electrophysiol. Published 13 February 2026. doi:10.1016/j.jacep.2025.12.039

Composed by AI from the published abstract and checked against it. How Briefs are made →