Coronary disease & ACS

Accuracy of instantaneous wave-free ratio and fractional flow reserve derived from single coronary angiographic projections

Am Heart JPublished Diagnostic accuracy study

Authors' conclusion

Angio-iFR did not achieve prespecified diagnostic performance against pressure wire-based iFR.

Study context

Population
Patients with coronary artery lesions of 40%–90% severity with iFR and fractional flow reserve (FFR) measurements; 32 centres in Europe, Japan and the US398 vessels paired
Participants
441 patients
Comparison
Angio-iFR vs pressure wire-based iFR
Outcome
Sensitivity and specificity of Angio-iFR vs invasively measured iFR (performance goals: sensitivity 75%, specificity 80%)
Methodological detailsprospective, multicentre registry, blinded independent corelab analysis, prespecified performance goals
  • Prospective
  • Multicentre registry
  • Blinded independent corelab analysis
  • Prespecified performance goals

Performance

As reported in the abstract

Per-vessel diagnostic performance vs wire-iFR
Test or modelSensitivitySpecificity
Angio-iFR77% (95% CI: 69%-84%)49% (95% CI: 41%-54%)

Angio-iFR

Sensitivity77% (95% CI: 69%-84%)
Specificity49% (95% CI: 41%-54%)

Further findings

Invasive iFR distribution (paired vessels)

Mean iFR (SD 0.11)
0.90
Vessels with iFR≤0.89
31.9%

Angio-iFR feasibility and analysis time

Feasibility
97%
Median analysis time
55 s

Original publication

Accuracy of instantaneous wave-free ratio and fractional flow reserve derived from single coronary angiographic projections. Am Heart J. Published 11 March 2025. doi:10.1016/j.ahj.2025.03.001

Summarised from the published abstract; every number is checked against it