Coronary disease & ACS

Accuracy of IVUS-Based Machine Learning Segmentation Assessment of Coronary Artery Dimensions and Balloon Sizing

JACC AdvPublished Diagnostic accuracy study

Authors' conclusion

ML model IVUS segmentation measurements were well-correlated with those of experts and selected an appropriate balloon size in more than 90% of images.

Study context

Population
Patients with intravascular ultrasound (IVUS) images of coronary arteries (IVUS-guided PCI optimization)test set: 437 images from 92 patients
Comparison
Machine learning (ML) automatic IVUS segmentation vs expert analysis
Outcome
Agreement rate between ML vs experts' measurements for appropriate balloon size selection, and lumen and acute stent areas (lumen area difference ≥0.5 mm² clinically significant)
Methodological detailsModel developed on 8,076 IVUS images from 234 patients, randomly split into training (83%) and validation (17%); evaluated on independent test set
  • Model developed on 8,076 IVUS images from 234 patients, randomly split into training (83%) and validation (17%); evaluated on independent test set

Performance

As reported in the abstract

Agreement rate ML vs experts (test data set)
Test or modelAgreement, lumen areaAgreement, acute stent areaAgreement, balloon size by vessel diameter onlyAgreement, balloon size adding lumen diameter
ML segmentation85.5%97.0%70.6%92.4%

ML segmentation

Agreement, lumen area85.5%
Agreement, acute stent area97.0%
Agreement, balloon size by vessel diameter only70.6%
Agreement, balloon size adding lumen diameter92.4%

Further findings

OutcomeML segmentation
Correlation with experts' segmentation (training data set)
Correlation coefficient, lumen area0.992
Correlation coefficient, vessel area0.993
  • Correlation with experts' segmentation (training data set): training set: 8,076 images from 234 patients

Original publication

Accuracy of IVUS-Based Machine Learning Segmentation Assessment of Coronary Artery Dimensions and Balloon Sizing. JACC Adv. Published 22 August 2023. doi:10.1016/j.jacadv.2023.100564

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.