Arrhythmias & electrophysiology

Accuracy of Smartwatches in the Detection of Atrial Fibrillation: A Systematic Review and Diagnostic Meta-Analysis

JACC AdvPublished Systematic review and meta-analysis of diagnostic accuracy studies

An erratum to this article was published (JACC Adv, May 2026 ↗)

Authors' conclusion

Smartwatches possess excellent diagnostic accuracy for AF detection

Study context

Population
Patients assessed for atrial fibrillation (AF) with smartwatches; studies published until January 2025
Participants
17,349 patients · 26 studies
Comparison
Smartwatches for AF detection
Outcome
Sensitivity, specificity and AUC of smartwatches for AF detection

Pooled result

As reported in the abstract

Overall diagnostic accuracy (all smartwatches)

Overall sensitivity

95%

95% CI 92%–97%; I2 = 95%

Overall specificity

97%

95% CI 94%–98%; I2 = 99%

Pooled AUC

0.97

95% CI 0.96%–0.99%

Further pooled results

OutcomeApple WatchSamsungWithings Scan Watch
Accuracy by device
Sensitivity94% (95% CI: 89%-96%; I2 = 71%)97% (95% CI: 92%-99%; I2 = 92%)89% (95% CI: 75%-96%; I2 = 94%)
Specificity97% (95% CI: 93%-99%; I2 = 87%)96% (95% CI: 91%-98%; I2 = 92%)95% (95% CI: 88%-99%; I2 = 93%)

Photoplethysmography vs ECG-based models

Accuracy comparable between photoplethysmography and ECG-based models

Original publication

Accuracy of Smartwatches in the Detection of Atrial Fibrillation: A Systematic Review and Diagnostic Meta-Analysis. JACC Adv. Published 3 November 2025. doi:10.1016/j.jacadv.2025.102133

Summarised from the published abstract; every number is checked against it

Get them every week

New studies on this topic in My Week and in your Monday e-mail.