Valvular & structural heart diseaseRandomised controlled trial

ACURATE neo2 valve versus commercially available transcatheter heart valves in patients with severe aortic stenosis (ACURATE IDE): a multicentre, randomised, controlled, non-inferiority trial

LancetPublished

  • Randomised controlled trial
  • multicentre
  • non-inferiority
  • Bayesian primary analysis

Authors' conclusion

In patients with symptomatic severe aortic stenosis, TAVR with ACURATE neo2 did not meet non-inferiority and resulted in significantly worse outcomes with respect to the primary endpoint of composite of all-cause mortality, all stroke, and rehospitalisation at 1 year when compared with commercial valves.

Study context

Population
Patients with symptomatic severe AS and any level of surgical risk, 71 centres in the US and Canada
Participants
1,500 patients
Comparison
Transcatheter aortic valve replacement (TAVR) with ACURATE neo2 vs commercially available valves
  • ACURATE neo2n = 752
  • Controln = 748reference
Primary endpoint
Composite of:
  • All-cause mortality
  • All stroke
  • Rehospitalisation (non-inferiority margin 8.0%)
Follow-up
1 year

Primary result

As reported in the abstract

Composite of:
  • All-cause mortality
  • Stroke
  • Rehospitalisation (posterior median probability)

Between-group difference

6.6%

95% CrI 3.0–10.2

By group

ACURATE neo2 vs Control
between-group difference 6.6% (95% CrI 3.0–10.2)

Non-inferiority margin 8 as printed, not drawn

Upper bound exceeded non-inferiority margin of 8%; non-inferiority not met; posterior probability of treatment difference
>0.999

Further findings

OutcomeACURATE neo2ControlHR (95% CI)P value
Composite endpoint, Kaplan-Meier rates · 1 yearKaplan-Meier rate 14.8% (95% CI 12.5–17.6)Kaplan-Meier rate 9.1% (95% CI 7.2–11.4)1.71 (1.26–2.33)P = 0.0005
All-cause mortality · 1 year36 events28 events1.30 (0.80–2.14)
Stroke · 1 year41 events25 events1.68 (1.02–2.75)
Rehospitalisation · 1 year38 events25 events1.57 (0.95–2.61)

CV mortality

P = 0.024

Safety

Spontaneous MI · 1 year

P = 0.0092

Prosthetic valve aortic regurgitation (central plus paravalvular) · 1 year

Mild

Moderate

Severe

  • Prosthetic valve aortic regurgitation (central plus paravalvular): mild p<0.0001
  • Prosthetic valve aortic regurgitation (central plus paravalvular): moderate p=0.0070
  • Prosthetic valve aortic regurgitation (central plus paravalvular): severe p=0.12

Original publication

ACURATE neo2 valve versus commercially available transcatheter heart valves in patients with severe aortic stenosis (ACURATE IDE): a multicentre, randomised, controlled, non-inferiority trial. Lancet. Published 21 May 2025. doi:10.1016/s0140-6736(25)00319-8

Summarised from the published abstract; every number is checked against it