Acute cardiac care & resuscitation

Adherence to automated external defibrillator instructions during real-life in-hospital cardiac arrest: an analysis of device data

ResuscitationPublished retrospective observational study

Authors' conclusion

Adherence to core AED prompts was high, whereas interruptive and corrective instructions were disregarded more frequently. AEDs for professionals should incorporate pre-analysis countdowns or alert tones to enhance user compliance. To ensure sustained high-quality compressions, devices should provide motivational-corrective instructions and prompt timely rescuer switches.

Study context

Population
Professionals responding to in-hospital cardiac arrest with automated external defibrillators (AEDs), real AED data from a German hospital between 2007 and 2024cases analyzed among 466 AED files
Participants
419 professionals

Main finding

As reported in the abstract

Prompt adherence

Prompt adherence ranged

15.3% to 81.1%

"start CPR" followed

81.1%

"press shock button" followed

77.7%

"push harder" followed

15.3%

Observational data: associations, not effects.

Further findings

Mean chest compression (CC) depth after the "push harder" prompt

Mean CC depth increased
from 3.26 cm to 3.66 cm, 3.61 cm, 3.51 cm
  • p < 0.001

Mean CC rate · first interval

Mean CC rate improved only within the first interval
from 89.8 cpm to 102.2 cpm
  • p < 0.001

High-quality CC with adherence to "push harder"

Adherence to "push harder" associated with high-quality CC
54.6% of cases
  • p < 0.001

Original publication

Adherence to automated external defibrillator instructions during real-life in-hospital cardiac arrest: an analysis of device data. Resuscitation. Published 29 January 2026. doi:10.1016/j.resuscitation.2026.110993

Summarised from the published abstract; every number is checked against it