Coronary disease & ACSDiagnostic or prognostic study

Anonymous Comparison of Various Angiography-Derived Fractional Flow Reserve Software With Pressure-Derived Physiological Assessment

JACC Cardiovasc IntervPublished

  • Diagnostic accuracy study
  • prospective cohort
  • independent core lab
  • blinded analysts
  • anonymized results

Authors' conclusion

This head-to-head comparison by an independent core lab demonstrated that the diagnostic accuracy of various angio-FFR software for predicting PW-FFR ≤0.80 was useful, with a higher discrimination compared with 2-dimensional QCA %DS; however, it did not reach the diagnostic accuracy previously reported in validation studies of various vendors.

Study context

Population
Vessels with documented sites of pressure wire-derived FFR and instantaneous wave-free ratio
Participants
390 vessels
Comparison
5 angio-FFR software/methods vs 2-dimensional quantitative coronary angiography (QCA) %DS, against PW-FFR
Outcome
AUC for predicting FFR ≤0.8

Performance

As reported in the abstract

Test or modelAUC
A0.75
B0.74
C0.74
D0.73
E0.73
2D QCA %DS0.65

Further findings

OutcomeValues
Analyzable vesselsA 100% · B 100% · C 92.1% · D 99.5% · E 92.1%

Original publication

Anonymous Comparison of Various Angiography-Derived Fractional Flow Reserve Software With Pressure-Derived Physiological Assessment. JACC Cardiovasc Interv. Published 16 May 2023. doi:10.1016/j.jcin.2023.04.026

Summarised from the published abstract; every number is checked against it