Cohort study
Association of the Timing and Extent of Cardiac Implantable Electronic Device Infections With Mortality
- Prospective cohort study
- multicentre
- 28 centres in Canada and the Netherlands
Authors' conclusion
Early systemic infections and delayed localized infections are associated with increased mortality, with the highest risk for patients with delayed systemic infections.
Study context
- Population
- Patients undergoing cardiac implantable electronic device (CIED) procedures, 28 centres in Canada and the Netherlands, 2012–2016177 developed infection
- Participants
- 19,559 patients
- Exposure
- CIED infection by timing and extent vs no CIED infection
- Early localizedn = 74
- Early systemicn = 45
- Delayed localizedn = 34
- Delayed systemicn = 23
- Outcome
- All-cause mortality by timing and extent of CIED infection (time-dependent analysis)
Main finding
As reported in the abstract
All-cause mortality
Each group vs no infectionAdjusted hazard ratio (95% CI)P value
Early localized
0.64(0.20–1.98)
P = 0.43Early systemic
2.88(1.48–5.61)
P = 0.002Delayed localized
3.57(1.33–9.57)
P = 0.01Delayed systemic
9.30(3.82–22.65)
P < 0.001Observational data: associations, not effects.
Further findings
Cumulative incidence of CIED infection · 3, 6 and 12 months
- Within 3 months
- 0.6%
- Within 6 months
- 0.7%
- Within 12 months
- 0.9%
- Infection rate in first 3 months (highest, reducing significantly thereafter)
- 0.21% per month
Original publication
Association of the Timing and Extent of Cardiac Implantable Electronic Device Infections With Mortality. JAMA Cardiol. Published 1 May 2023. doi:10.1001/jamacardio.2023.0467