Valvular & structural heart diseaseCohort study

Asymmetrical Expansion of Balloon-Expandable Transcatheter Aortic Valve Prostheses

JACC Cardiovasc IntervPublished

  • Prospective registry study

Authors' conclusion

Asymmetrical expansion of balloon-expandable THVs was associated with impaired hemodynamic valve performance but not with clinical outcomes.

Study context

Population
Patients undergoing transfemoral transcatheter aortic valve replacement (TAVR) for native severe AS with contemporary balloon-expandable transcatheter heart valves (THVs), February 2014 to June 2022
Participants
1,216 patients
Exposure
TAVR asymmetry index
Outcome
Impaired hemodynamic valve performance (mean residual THV gradient ≥20 mm Hg and/or moderate or greater paravalvular regurgitation) on echocardiography
Follow-up
Before hospital discharge

Main finding

As reported in the abstract

Asymmetry index as predictor of impaired hemodynamic valve performance

AUC of asymmetry index

0.88 (95% CI: 0.84-0.92; P < 0.001)

Optimal threshold

>5.5% (sensitivity 77%, specificity 86%)

Patients above threshold

17%

Higher asymmetry index (continuous), adjusted OR

1.37 (95% CI: 1.29-1.46; P < 0.001)

Observational data: associations, not effects.

Further findings

All-cause mortality, high TAVR asymmetry index · median 376 days of follow-up

ComparisonHR (95% CI)P value
High asymmetry index1.00 (0.71–1.41)P = 0.989

Original publication

Asymmetrical Expansion of Balloon-Expandable Transcatheter Aortic Valve Prostheses. JACC Cardiovasc Interv. Published 9 September 2024. doi:10.1016/j.jcin.2024.05.039

Summarised from the published abstract; every number is checked against it