Atrioventricular synchrony maintained by a dual-chamber leadless pacemaker in real-world conditions

EuropacePublished

  • Prospective single-arm clinical study
  • multicentre
  • independent Holter core laboratory

Authors' conclusion

A dual-chamber leadless pacing system demonstrated >96% overall AV synchrony for a 24 h period of daily living while programmed to clinically appropriate settings, thus showing that true dual-chamber DDD(R) pacing can be maintained in a real-world environment.

Study context

Population
Patients with standard indications for dual-chamber pacing, 24 h Holter after the 3-month visit during daily living
Participants
47 analysable patients
Outcome
AV synchrony (proportion of synchronous beats, PR intervals within paced and sensed AV delay bounds)
Follow-up
24 h period of daily living after the 3-month visit

Main finding

As reported in the abstract

AV synchrony (proportion of synchronous beats, Holter core laboratory adjudicated) · 24 h ambulatory period after the 3-month visit

Mean atrioventricular (AV) synchrony (% of beats)

96.7%

Mean A-to-V transmission success rate

93.4%

Mean V-to-A transmission success rate

92.1%

Observational data: associations, not effects.

Further findings

AV synchrony by implant indication (sinus node vs AV block)

No significant differences in AV synchrony between sinus node and AV block patients

AV synchrony across AV paced/sensed cycle combinations and heart rate ranges (including >100 b.p.m.)

AV synchrony sustained across all combinations and heart rate ranges
>90%

Original publication

Atrioventricular synchrony maintained by a dual-chamber leadless pacemaker in real-world conditions. Europace. Published 19 January 2026. doi:10.1093/europace/euag012

Summarised from the published abstract; every number is checked against it