Heart failure & cardiomyopathies

Automatic measurements of left ventricular volumes and ejection fraction by artificial intelligence: clinical validation in real time and large databases

Eur Heart J Cardiovasc ImagingPublished Validation study

Performance · Median 5.3 minResult — go to the resultReduction 77%

Authors' conclusion

The software reduced the time taken to perform and volumetrically analyse routine echocardiograms without a decrease in accuracy compared with experts.

Study context

Population
Patients undergoing echocardiography: 50 consecutive patients (real-time), 40 subjects with repeated exams, internal and external research databases of 1881 and 849 subjects
Comparison
Artificial intelligence (AI) automatic left ventricular (LV) volume and ejection fraction (EF) measurements vs manual measurements
Outcome
Acquisition and processing time, test-retest reproducibility and agreement with reference of LV volumes and EF

Performance

Median 5.3 min

Reduction

77%

P < 0.001

As reported in the abstract

Original publication

Automatic measurements of left ventricular volumes and ejection fraction by artificial intelligence: clinical validation in real time and large databases. Eur Heart J Cardiovasc Imaging. Published 26 October 2023. doi:10.1093/ehjci/jead280

Composed by AI from the published abstract and checked against it. How Briefs are made →