Congenital heart diseaseRandomised controlled trial

Bioresorbable vs Metallic Occluders for Transcatheter Atrial Septal Defect Closure

JAMAPublished

  • Randomised controlled trial
  • multicentre
  • non-inferiority
  • open-label
  • 10 hospital sites in China

Authors' conclusion

A bioresorbable occluder is noninferior to a metallic occluder for ASD closure, with near-complete degradation by 2 years.

Study context

Population
Participants with secundum atrial septal defect (ASD)229 implanted
Participants
230 participants
Comparison
Bioresorbable vs metallic occluder for transcatheter ASD closure
  • Bioresorbable occludern = 116
  • Metallic occludern = 114reference
Primary endpoint
ASD closure success (procedural success with residual shunt ≤2 mm on transthoracic echocardiography)
Follow-up
6 months

Primary result

As reported in the abstract

Between-group difference

−0.8 percentage points

95% CI −5.0 to 3.7

P < 0.001 for non-inferiority

Events by treatment group

Bioresorbable occluder vs Metallic occluder
between-group difference −0.8 percentage points (95% CI −5.0 to 3.7); P for non-inferiority < 0.001

Further findings

ASD closure success · 2 years

P = 0.75

Degradation of bioresorbable occluder · 2 years
99.8%

Safety

Device-related adverse events · 2 years

P = 0.72

Original publication

Bioresorbable vs Metallic Occluders for Transcatheter Atrial Septal Defect Closure. JAMA. Published 23 October 2025. doi:10.1001/jama.2025.17639

Summarised from the published abstract; every number is checked against it