Valvular & structural heart diseaseRandomised controlled trial

Cardiac Magnetic Resonance Imaging Versus Computed Tomography to Guide Transcatheter Aortic Valve Replacement: A Randomized, Open-Label, Noninferiority Trial

CirculationPublished

  • Randomised controlled trial
  • investigator-initiated
  • prospective
  • open-label
  • non-inferiority
  • 2 Austrian heart centres

Authors' conclusion

CMR-guided TAVR was noninferior to CT-guided TAVR in terms of device implantation success.

Study context

Population
Transcatheter aortic valve replacement (TAVR) candidates with severe symptomatic AS, excluding CKD level 4 or 5267 underwent TAVR (mITT)
Participants
380 candidates
Comparison
CMR-guided vs computed tomography (CT)-guided TAVR planning
  • CMRn = 191
  • CTn = 189reference
Primary endpoint
VARC-2 implantation success (absence of procedural mortality, correct positioning of a single prosthetic valve, proper prosthetic valve performance)
Follow-up
Discharge

Primary result

Implantation success (modified ITT cohort)

Between-group difference

2.8%

90% CI −2.7% to 8.2%

P < 0.01 for non-inferiority

Events by treatment group

CMR vs CT
between-group difference 2.8% (90% CI −2.7% to 8.2%); P for non-inferiority < 0.01

As reported in the abstract

Further findings

Implantation success, per-protocol cohort · discharge

Between-group difference, per-protocol cohort (n=248)
2.0% (90% CI, −3.8% to 7.8%; P<0.01 for noninferiority)
  • per-protocol cohort of 248 patients (121 CMR-guided, 127 CT-guided)
  • non-inferiority margin 9% (absolute risk difference)

Original publication

Cardiac Magnetic Resonance Imaging Versus Computed Tomography to Guide Transcatheter Aortic Valve Replacement: A Randomized, Open-Label, Noninferiority Trial. Circulation. Published 27 August 2023. doi:10.1161/circulationaha.123.066498

Summarised from the published abstract; every number is checked against it