Coronary disease & ACSDiagnostic or prognostic study

Clinical and Analytical Performance of a Novel Point-of-Care High-Sensitivity Cardiac Troponin I Assay

J Am Coll CardiolPublished

  • Diagnostic accuracy study
  • international
  • multicentre

Final diagnosis centrally adjudicated by 2 independent cardiologists

Authors' conclusion

The SPINCHIP hs-cTnI POC test has very high diagnostic accuracy. Its assay-specific 0/1-hour algorithm achieved very high sensitivity/negative predictive value and specificity/positive predictive value for rule-out/in MI.

Study context

Population
Adults presenting to the ED with acute chest discomfortMyocardial infarction (MI) in 214 (19%)
Participants
1,102 adults
Comparison
hs-cTnI-SPINCHIP point-of-care (POC) vs central laboratory assays
Outcome
Diagnostic performance (AUC) for centrally adjudicated MI

Performance

As reported in the abstract

AUC for MI
Test or modelAUC
hs-cTnI-SPINCHIP0.94 (95% CI: 0.92-0.95)
hs-cTnI-Architect0.94 (95% CI: 0.92-0.95)
hs-cTnT Elecsys0.93 (95% CI: 0.91-0.95)
  • hs-cTnI-SPINCHIP vs hs-cTnI-Architect: P = 0.907
  • hs-cTnI-SPINCHIP vs hs-cTnT Elecsys: P = 0.305

Further findings

OutcomeValues
hs-cTnI-SPINCHIP 0/1-h algorithmhs-cTnI-SPINCHIP Ruled out (<7 ng/L at presentation if onset >3 h, or <7 ng/L with 0/1-h delta <4 ng/L) 51% · Sensitivity 100% (95% CI: 97.7%-100%) · NPV 100% (95% CI: 99.0%-100%) · Ruled in (≥36 ng/L or 0/1-h delta ≥11 ng/L) 27% · Specificity 90.9% (95% CI: 88.3%-92.9%) · PPV 72.9% (95% CI: 66.4%-78.6%)

hs-cTnI-SPINCHIP 0/1-h algorithm

Bootstrap internal validation confirmed excellent diagnostic performance

Sample type comparison (whole blood, fresh/frozen plasma, capillary finger prick)

High agreement between sample types

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Clinical and Analytical Performance of a Novel Point-of-Care High-Sensitivity Cardiac Troponin I Assay. J Am Coll Cardiol. Published 20 August 2024. doi:10.1016/j.jacc.2024.05.056

Summarised from the published abstract; every number is checked against it