Arrhythmias & electrophysiologyRandomised controlled trial

Comparison of Pacing Lead Design on Left Bundle Branch Pacing Outcomes

JACC Clin ElectrophysiolPublished

  • multicenter, blind-adjudicated, randomized clinical trial
  • block randomization in a 1:1 ratio

Authors' conclusion

Among LBBP recipients, SDL compared with LLL resulted in a higher incidence of LLC during follow-up, lower acute LBBP procedural success, and greater acute lead damage during implantation. These findings have significant implications for lead selection in LBBP.(Comparison of Pacing Lead Design on Left Bundle Branch Pacing Outcomes: The LEAD_LBBP Randomized Clinical Trial [LEAD-LBBP]; NCT06318130).

Study context

Population
Left bundle branch pacing (LBBP) recipientsmean age 74.9 ± 9.6 years, 97 [44%] female
Participants
226 recipients
Comparison
LLL or SDL
  • LLLn = 113reference
  • SDLn = 113
Primary endpoint
Incidence of LLC
Follow-up
12 months

Primary result

As reported in the abstract

Hazard ratio

2.72

95% CI 1.34–5.54

P = 0.006

By group

SDL vs LLL
HR 2.72 (95% CI 1.34–5.54); P = 0.006
LLC overall
36 (16%)

Secondary outcomes

Acute LBBP success

  • Acute LBBP success: P = 0.007

Pacing parameters and clinical outcomes

Comparable between the SDL and LLL groups
P > 0.05

Safety

OutcomeSDLLLLOR (95% CI)P value
Acute lead damage10%3%3.95 (1.07–14.58)P = 0.027

Overall complications

Comparable between the SDL and LLL groups
P > 0.05

Original publication

Comparison of Pacing Lead Design on Left Bundle Branch Pacing Outcomes. JACC Clin Electrophysiol. Published 27 June 2026. doi:10.1016/j.jacep.2026.06.008

Summarised from the published abstract; every number is checked against it