Arrhythmias & electrophysiologyCohort study

Comparison of Patient Outcomes Between Leadless vs Transvenous Pacemakers Following Transcatheter Aortic Valve Replacement

JACC Cardiovasc IntervPublished

  • Retrospective cohort study
  • propensity score overlap weighting

Authors' conclusion

Leadless pacemakers are increasingly being used for conduction disturbance following TAVR and were associated with a lower rate of in-hospital complications and midterm device-related complications compared to transvenous pacemakers without a difference in midterm mortality.

Study context

Population
Medicare fee-for-service beneficiaries aged ≥65 years with pacemaker following transcatheter aortic valve replacement (TAVR), 2017–2020
Participants
10,338 beneficiaries
Exposure
Leadless vs transvenous pacemakers after TAVR
  • Leadlessn = 730
  • Transvenousn = 9,608reference

Main finding

As reported in the abstract

In-hospital overall complications · in-hospital

P = 0.014

Observational data: associations, not effects.

Further findings

OutcomeEstimateP value
All-cause death · midterm (up to 2 years)adjusted HR 1.13 (95% CI 0.96–1.32)P = 0.15
HF hospitalization · midterm (up to 2 years)subdistribution HR 0.89 (95% CI 0.74–1.08)P = 0.24
Infective endocarditis · midterm (up to 2 years)subdistribution HR 0.98 (95% CI 0.44–2.17)P = 0.95

Trend in leadless pacemaker use after TAVR, 2017–2020

Increase in proportion of leadless pacemakers
3.5-fold

Safety

Outcomesubdistribution HR (95% CI)P value
Device-related complications · midterm (up to 2 years)0.37 (0.21–0.64)P < 0.001

Original publication

Comparison of Patient Outcomes Between Leadless vs Transvenous Pacemakers Following Transcatheter Aortic Valve Replacement. JACC Cardiovasc Interv. Published 17 July 2024. doi:10.1016/j.jcin.2024.05.030