Arrhythmias & electrophysiologyCohort study

Comprehensive analysis of substernal lead removal: experience from EV ICD Pilot, Pivotal, and Continued Access Studies

EuropacePublished

  • Retrospective analysis of the EV ICD Pilot, Pivotal and Continued Access Studies
  • pooled 3 studies
  • Mean lead dwell time 12.6 ± 14.3 months (0.2-58.4)

Authors' conclusion

Complete removal of the EV ICD lead was successful in 93.1% of cases, and simple traction was sufficient in most instances.

Study context

Population
Patients with a successful EV ICD implant who underwent lead removalof 347 implanted
Participants
29 patients
Outcome
Lead removal success

Main finding

As reported in the abstract

Lead removal successful

27/29 (93.1%)

Success <1 year post-implant

100% (19/19)

Success >1 year post-implant

80% (8/10)

Observational data: associations, not effects.

Further findings

Removal technique (successful cases with known technique)

Simple traction
22/26 (84.6%)
Extraction tools
4/26 (15.4%)

EV ICD reimplantation

Reimplant attempts successful
11
  • All 11 attempts successful

Reason for removal

Lead dislodgement (main reason)
n = 9, 31.0%

Safety

Removal procedure complications

No complications reported

Original publication

Comprehensive analysis of substernal lead removal: experience from EV ICD Pilot, Pivotal, and Continued Access Studies. Europace. Published 30 August 2024. doi:10.1093/europace/euae225