Arrhythmias & electrophysiologyCohort study

Cryocure-VT: the safety and effectiveness of ultra-low-temperature cryoablation of monomorphic ventricular tachycardia in patients with ischaemic and non-ischaemic cardiomyopathies

EuropacePublished

  • Single-arm clinical trial
  • prospective
  • multicentre
  • first-in-human
  • 9 centres

Authors' conclusion

In this first-in-human multi-centre experience, endocardial ULTC ablation of monomorphic VT appears safe and effective in patients with both ischaemic-cardiomyopathy and non-ischaemic-cardiomyopathy.

Study context

Population
Patients referred for de novo or second ablation of recurrent monomorphic VT, ischaemic and non-ischaemic aetiologiesage 67 ± 11 years, 78% ischaemic
Participants
64 patients
Outcome
Freedom from device- or procedure-related MAEs; acute non-inducibility of clinical VTs at end of procedure; freedom from sustained VT or ICD intervention
Follow-up
30 days (MAEs); end of procedure; 6 months

Main finding

As reported in the abstract

Acute non-inducibility of clinical ventricular tachycardias (VTs) (patients with post-ablation induction attempted)

94% (51/54)

Protocol-defined major adverse events (MAEs) up to 30 days

0

VT-free at 6 months

38 patients (60.3%)

Freedom from defibrillator shock at 6 months

81.0%

Observational data: associations, not effects.

Further findings

VT burden before vs after ablation (patients with ICD ≥6 months before ablation, n = 47)

Before ablation, median (IQR)
4 (1-9)
After ablation, median (IQR)
0 (0-2)

Safety

Procedure-related serious adverse events

Procedure-related serious adverse events, resolved without clinical sequelae
4
Protocol-defined MAEs
0

Original publication

Cryocure-VT: the safety and effectiveness of ultra-low-temperature cryoablation of monomorphic ventricular tachycardia in patients with ischaemic and non-ischaemic cardiomyopathies. Europace. Published 7 April 2024. doi:10.1093/europace/euae076

Summarised from the published abstract; every number is checked against it