Coronary disease & ACS

Deep Learning and Fluid Dynamics On-Site CT-FFR Solution Compared to Off-Site FFRct and Invasive Fractional Flow Reserve (FFR)

JACC Cardiovasc ImagingPublished Diagnostic accuracy study

Performance · Diagnostic performance for functionally significant CADResult — go to the resultAUC 0.91 vs 0.89

Authors' conclusion

This study establishes xFFR as a robust and efficient on-site tool for assessing CAD, demonstrating high diagnostic accuracy, reproducibility, and agreement with invasive methods.

Study context

Population
Symptomatic patients at intermediate-to-high coronary artery disease (CAD) risk undergoing coronary computed tomography angiography (CTA) and invasive coronary angiography
Participants
250 patients
Comparison
On-site xFFR vs off-site FFRct, with invasive FFR as reference
Outcome
Diagnostic performance for functionally significant CAD vs iFFR (AUC)
Methodological detailssingle-centre, prospective
  • Single-centre
  • Prospective

Performance

As reported in the abstract

Diagnostic performance for functionally significant CAD

AUC

0.91 vs 0.89

P = 0.274

By group

Further findings

OutcomexFFRFFRctiFFRAUCP value
AUC, left anterior descending artery0.96 vs 0.84P = 0.001
AUC0.960.84Not reported in the abstractiFFR: not reported in the abstract
xFFR analysis time
8 · mean · SD 3.4 · minutes

Correlation of xFFR (Spearman)

xFFR vs iFFR
ρ = 0.67
xFFR vs FFRct
ρ = 0.53

Original publication

Deep Learning and Fluid Dynamics On-Site CT-FFR Solution Compared to Off-Site FFRct and Invasive FFR. JACC Cardiovasc Imaging. Published 27 February 2026. doi:10.1016/j.jcmg.2025.11.011

Summarised from the published abstract; every number is checked against it

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.