Acute cardiac care & resuscitation
Device-Related Adverse Events and Outcomes in Patients With Temporary Mechanical Circulatory Support Placed at Referral Centers Versus Cardiogenic Shock Hub Centers: An Observational Analysis
- Observational Analysis
- single center
+1 more
- single center
Authors' conclusion
Initial tMCS placement for CS at an RRC with transfer to an HC is associated with a higher DRAE prevalence and worse outcomes than initial tMCS placement at a CS HC. The higher mortality in RRC-implanted patients may be due to greater CS severity.
Study context
- Population
- Patients with temporary mechanical circulatory support (temporary mechanical support (tMCS)) for cardiogenic shock (CS), August 2021 to August 2024, stratified by location of initial tMCS device placement77% HC-implanted, 23% RRC-implanted
- Participants
- 398 patients
- Exposure
- tMCS placement for CS at an RRC before transfer to an HC compared with initial tMCS device placement at an HC
Main finding
As reported in the abstract
Unadjusted in-hospital mortality
Each group vs HC-implantedOdds ratio (95% CI)P value
RRC-implantedunadjusted
2.52(1.52–4.18)
P < 0.001In-hospital mortality after adjustment for CS severity and cardiac arrest — RRC-implanted
1.72(0.95–3.12)
P = 0.07Observational data: associations, not effects.
Further findings
| Outcome | odds ratio (95% CI) | P value | |
|---|---|---|---|
| Unfavorable outcomes (death before discharge, heart transplant, or durable left ventricular assist device) — RRC-implanted vs HC-implanted · unadjusted | 2.55 (1.52–4.27) | P < 0.001 | |
| Unfavorable outcomes (death before discharge, heart transplant, or durable left ventricular assist device) — RRC-implanted vs HC-implanted · after adjustment for CS severity and cardiac arrest | 1.60 (0.87–2.92) | P = 0.13 | |
| 0.20.5125Odds ratio (log scale) |
Baseline characteristics
- RRC-implanted patients more commonly experienced cardiac arrest and had more advanced CS
Safety
DRAE prevalence
any DRAE
bleeding
hemolysis
vascular injury
DRAE rate · numerically higher among RRC-implanted
- DRAE rate — Overall DRAE rate
- 0.33 events/patient-week
- DRAE prevalence: P <0.05 for all
Original publication
Device-Related Adverse Events and Outcomes in Patients With Temporary Mechanical Circulatory Support Placed at Referral Centers Versus Cardiogenic Shock Hub Centers: An Observational Analysis. Circ Heart Fail. Published 13 April 2026. doi:10.1161/circheartfailure.125.013742
Summarised from the published abstract; every number is checked against it