Arrhythmias & electrophysiologyCohort study
Durability of pulmonary vein isolation: Does the pulsed field ablation system matter?
- retrospective review
- consecutive patients
- 2 centers
- propensity score-matched sensitivity analysis
Authors' conclusion
The balloon-in-basket PFA platform demonstrated significantly greater chronic PVI durability than the pentaspline system. Propensity score matching confirmed that this durability advantage was not attributable to baseline substrate differences. These platform-specific differences are clinically meaningful and warrant prospective validation.
Study context
- Population
- Patients undergoing repeat ablation after a de novo pulsed field ablation (PFA) procedure for atrial fibrillation, between May 2023 and September 2025de novo PFA by catheter, 17 underwent repeat procedures, propensity score-matched n = 34 pairs · pentaspline 86 · balloon-in-basket 46
- Participants
- 86 vs 46
- Exposure
- chronic pulmonary vein isolation (PVI) durability and pulmonary vein (PV) reconnection patterns after index ablation with FARAPULSE (pentaspline) and Volt (balloon-in-basket) catheters
- pentasplinen = 86
- balloon-in-basketn = 46
Main finding
As reported in the abstract
patients with clinical recurrence (9 in each group)
18
of whom underwent repeat procedures (balloon-in-basket n = 9; pentaspline n = 8)
17
Observational data: associations, not effects.
Further findings
PV-level reconnection
Patient-level durability
Clinical recurrence rate · numerically higher but nonsignificant
Clinical recurrence rate, propensity-matched cohort
- PV-level reconnection: P = .003
- Patient-level durability: P = .049
- Clinical recurrence rate: P = .12
- Clinical recurrence rate, propensity-matched cohort: P = 1.000
- Clinical recurrence rate, propensity-matched cohort: propensity score-matched sensitivity analysis (n = 34 pairs)
Arrhythmia recurrence by posterior wall isolation, combined cohort
- with posterior wall isolation vs without
- 4.3% vs 17.6%
- P = .031
Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.
Original publication
Summarised from the published abstract; every number is checked against it