Valvular & structural heart disease

Early safety and feasibility of a first-in-class biomimetic transcatheter aortic valve - DurAVR

EuroInterventionPublished First-in-human prospective single-arm study

Authors' conclusion

Preliminary results from the FIH study with DurAVR THV demonstrate a good safety profile with promising haemodynamic performance sustained at 1 year and restoration of near-normal flow dynamics.

Study context

Population
Patients with severe symptomatic aortic stenosis (AS) of any surgical risk eligible for DurAVR transcatheter heart valve (THV)
Participants
13 patients
Outcome
Safety and feasibility (implant success, haemodynamic performance, safety)
Follow-up
30 days, 6 months, 1 year
Methodological detailsnon-randomised, single-centre
  • Non-randomised
  • Single-centre

Main finding

As reported in the abstract

Implant success

DurAVR THV

100%

Observational data: associations, not effects.

Further findings

Haemodynamics: EOA (mean annulus 22.95±1.09 mm) · 30 days

EOA
2.00±0.17 cm²
MPG
9.02±2.68 mmHg

Haemodynamics: EOA and MPG · 1 year

EOA
1.96±0.11 cm²
MPG
8.82±1.38 mmHg
Patients with any degree of prosthesis-patient mismatch
0

CMR valve performance

Mean coaptation length
8.3±1.7 mm
Restoration of laminar flow, consistent with a predisease state

Safety

Procedural and follow-up safety · up to 1 year

Implant success
100%
Device-related complications
0
Access site complication
1
Permanent pacemaker implantation
1
Moderate aortic regurgitation
1
Deaths, stroke, bleeding, reinterventions or MI
0

Original publication

Early safety and feasibility of a first-in-class biomimetic transcatheter aortic valve - DurAVR. EuroIntervention. Published 17 July 2023. doi:10.4244/eij-d-23-00282