Arrhythmias & electrophysiologyRandomised controlled trial

Efficacy and safety of a novel size-adjustable cryoballoon vs conventional cryoballoon: Results from the CONTRAST-CRYO II randomized clinical trial

Heart RhythmPublished

  • randomized clinical trial

Authors' conclusion

Contrary to our hypothesis, the PNI rate was significantly higher with SACB than with CCB, especially during right inferior PV ablation using the 31-mm balloon. Other safety outcomes and 1-year efficacy were comparable in groups.

Study context

Population
Patients undergoing pulmonary vein isolation for atrial fibrillationunderwent ablation according to the allocation · CCB 109 · size-adjustable cryoballoon (SACB) 110
Participants
219 patients
Comparison
SACB vs CCB
  • SACBn = 110
  • CCBn = 109
Primary endpoint
Incidence of phrenic nerve injury (PNI)

Primary result

As reported in the abstract

P = 0.02

Secondary outcomes

Cryoballoon PV isolation success

P = 0.42

Single-shot isolation rates

Arrhythmia-free rates · 1 year · log-rank

P = 0.74

Pulmonary vein (PVs) isolated using the 31-mm balloon in the SACB-group
87.3% · 378 of 433 PVs of 110

Procedural parameters

Largely comparable, except for significantly lower balloon temperatures with SACB
Procedure and fluoroscopy times were similar

Original publication

Efficacy and safety of a novel size-adjustable cryoballoon vs conventional cryoballoon: Results from the CONTRAST-CRYO II randomized clinical trial. Heart Rhythm. Published 26 May 2026. doi:10.1016/j.hrthm.2026.05.037

Summarised from the published abstract; every number is checked against it