Valvular & structural heart diseaseResearch article

Electrosurgical laceration and stabilisation of tricuspid edge-to-edge repair: the ELASTA-T technique

EuroInterventionPublished

  • step-by-step standardised description of the ELASTA-T strategy

Authors' conclusion

By removing any mechanical obstruction from centrally placed clips, the technique facilitates secure TTVR anchoring and may reduce paravalvular regurgitation. This step-by-step framework may support procedural standardisation and broader adoption, ultimately improving outcomes in this high-risk population.

Study context

Population
Patients with recurrent tricuspid valve regurgitation (TVR) after tricuspid transcatheter edge-to-edge repair (T-TEER), particularly when centrally positioned clips impede subsequent transcatheter tricuspid valve replacement (TTVR)

Main finding

As reported in the abstract

ELASTA-T technique
  • ELASTA-T allows safe and reproducible single leaflet device attachment (SLDA), creating adequate central space for accurate positioning and full expansion of a transcatheter tricuspid valve (TTV)
  • Detachment can be reliably confirmed by fluoroscopy and transoesophageal echocardiography

Original publication

Electrosurgical laceration and stabilisation of tricuspid edge-to-edge repair: the ELASTA-T technique. EuroIntervention. Published 20 April 2026. doi:10.4244/eij-d-25-01316

Summarised from the published abstract; every number is checked against it