Heart failure & cardiomyopathies

Enhancing Sweat Rate Using a Novel Device for the Treatment of Congestion in Heart Failure

Circ Heart FailPublished Feasibility study

Authors' conclusion

Enhancing sweat rate was safe and resulted in a clinically meaningful fluid removal and weight loss.

Study context

Population
HF patients with congestion and normal subjects3 sessions of up to 4 h each
Participants
6 vs 18
Comparison
Normal vs Patients with congestion
  • Normaln = 6
  • Patients with congestionn = 18
Outcome
Ability to increase skin temperature to the desired range (35 °C to 38 °C)

Main finding

As reported in the abstract

Skin temperature
  • Skin temperature increased to median 37.5 °C (IQR 37.1-37.9 °C)
  • Median core temperature increase 0.2 °C (IQR 0.1-0.3 °C)

Secondary outcomes

Hourly sweat output ≥150 mL/h

Treatment procedures with average sweat rate ≥150 mL/h
80%

Hourly weight loss during treatment

Median 215 g/h (IQR 165-285; range 100-344 g/h)

Safety

No procedure-related adverse events
No significant changes in hemodynamic variables or renal function

Original publication

Enhancing Sweat Rate Using a Novel Device for the Treatment of Congestion in Heart Failure. Circ Heart Fail. Published 2 November 2022. doi:10.1161/circheartfailure.122.009787