Arrhythmias & electrophysiology
Epicardial pacing outcomes in infants with heart block: Lead and device complications from a multicenter experience
Authors' conclusion
Device complications affect 15% of infants receiving a permanent PM for heart block. Age ≤90 days at PM implantation is especially associated with infectious complications. Epicardial lead durability appears similar to previously reported pediatric experiences.
Study context
- Population
- Infants with complete heart block receiving an epicardial pacemaker, 2000–2021282 bipolar, 39 unipolar leads
- Participants
- 174 infants
- Outcome
- Time to device-related adverse event:
- Lead failure requiring revision
- Pocket infection
- Exit block requiring increased pacing output
- Lead-related coronary artery compression
Methodological detailsmulticentre
- Multicentre
Main finding
As reported in the abstract
Device-related adverse event · median follow-up 2.1 years
Infants with primary outcome
26
Median time to event
0.6 year
Age ≤90 days at implantation, risk of device-related adverse event (most significant risk factor; mainly pocket infections), HR
7.02
P < .001
Observational data: associations, not effects.
Further findings
- Leads with failure
- 3%
- 5-year freedom from failure
- 93%
- 10-year freedom from failure
- 83%
Original publication
Epicardial pacing outcomes in infants with heart block: Lead and device complications from a multicenter experience. Heart Rhythm. Published 14 July 2024. doi:10. 1016/ j. hrthm. 2024. 07. 014