Arrhythmias & electrophysiology

Epicardial pacing outcomes in infants with heart block: Lead and device complications from a multicenter experience

Heart RhythmPublished Retrospective cohort study

Authors' conclusion

Device complications affect 15% of infants receiving a permanent PM for heart block. Age ≤90 days at PM implantation is especially associated with infectious complications. Epicardial lead durability appears similar to previously reported pediatric experiences.

Study context

Population
Infants with complete heart block receiving an epicardial pacemaker, 2000–2021282 bipolar, 39 unipolar leads
Participants
174 infants
Outcome
Time to device-related adverse event:
  • Lead failure requiring revision
  • Pocket infection
  • Exit block requiring increased pacing output
  • Lead-related coronary artery compression
Methodological detailsmulticentre
  • Multicentre

Main finding

As reported in the abstract

Device-related adverse event · median follow-up 2.1 years

Infants with primary outcome

26

Median time to event

0.6 year

Age ≤90 days at implantation, risk of device-related adverse event (most significant risk factor; mainly pocket infections), HR

7.02

P < .001

Observational data: associations, not effects.

Further findings

Leads with failure
3%
5-year freedom from failure
93%
10-year freedom from failure
83%

Original publication

Epicardial pacing outcomes in infants with heart block: Lead and device complications from a multicenter experience. Heart Rhythm. Published 14 July 2024. doi:10.1016/j.hrthm.2024.07.014

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.