Valvular & structural heart diseaseCohort study

Evaluation of tricuspid valve regurgitation following transvenous rotational mechanical lead extraction

EuropacePublished

  • Observational cohort study

Authors' conclusion

Significant TRI was detected in 5.7% of patients following transvenous rotational mechanical lead extraction. The duration of lead implantation emerged as the sole predictor of significant TRI.

Study context

Population
Patients undergoing mechanical rotational transvenous lead extraction (TLE)290 leads extracted
Participants
158 patients
Outcome
Rate of significant acute tricuspid regurgitation increase (TRI: increase ≥1 grade with post-extraction severity at least moderate) by echocardiography

Main finding

As reported in the abstract

Significant TRI

Significant TRI

5.7%

Observational data: associations, not effects.

Further findings

Univariate predictors of significant TRI

ComparisonOR (95% CI)P value
implant duration of all leads · univariate1.01 (1.003–1.018)P = 0.001
right ventricular leads · univariate; implant duration1.01 (1.004–1.017)P = 0.002

Independent predictors of mortality

ComparisonHR (95% CI)P value
Severe increase of TR following TLE · independent predictor5.20 (1.44–18.73)P = 0.012
severe systolic dysfunction2.37 (1.01–5.20)P = 0.032
systemic infection2.28 (1.06–4.89)P = 0.035

Original publication

Evaluation of tricuspid valve regurgitation following transvenous rotational mechanical lead extraction. Europace. Published 11 July 2024. doi:10.1093/europace/euae191

Summarised from the published abstract; every number is checked against it