Evolution of Tined Leadless Pacemaker Implantation: Patient Selection, Implant Volume, Center Characteristics, and Safety Trends

Heart RhythmPublished

  • Medicare claims linked to a manufacturer's device registration data
  • random-intercept models
+1 more
  • random-intercept models

Authors' conclusion

In Medicare patients with tined leadless pacemakers, cardiac perforation rates declined over time and were lower in those implanted by more experienced implanters. Variation in outcomes was more closely associated with implanters than facilities.

Study context

Population
Medicare patients implanted with a tined leadless pacemaker from 2017 to 2024implants, 19,446 (41.6%) were Micra AV
Participants
46,698 patients
Comparison
Cardiac perforation rates over time and by implanter volume and facility characteristics

Main finding

As reported in the abstract

Cardiac perforation rates over time · 2017 to 2024

Adjusted Coefficient

−0.13, 95% CI −0.19 - −0.06, p-value<0.01

Further findings

OutcomeOR (per additional 10 implants) (95% CI)P value
Cardiac perforation by implanter volume · higher implanter volume associated with fewer perforations0.93 (0.88–0.98)P < 0.01

Patient characteristics

Mean age
80.4 years
Charlson Comorbidity Index
4.7±3.4

Patient selection across providers and facilities

Differences in patient characteristics across providers and facilities were observed; adjustment did not materially alter the associations between implanters and facility and perforation rates

Original publication

Evolution of Tined Leadless Pacemaker Implantation: Patient Selection, Implant Volume, Center Characteristics, and Safety Trends. Heart Rhythm. Published 16 October 2026. doi:10.1016/j.hrthm.2026.10.014