Arrhythmias & electrophysiologyCohort study

Extraction outcomes of implantable cardioverter-defibrillator leads vary by manufacturer and model family

EuropacePublished

  • Prospective registry cohort study
  • single-centre
  • consecutive patients

Authors' conclusion

Complete and safe ICD lead removal rate by TLE is extremely high but varied by manufacturer and lead family. Linox family lead and >10 years lead age were independent predictors of incomplete lead removal.

Study context

Population
Patients with implantable cardioverter-defibrillator (ICD) undergoing transvenous lead extraction (TLE), Cleveland Clinic Prospective TLE Registry, 2013–2022885 ICD leads
Participants
810 patients
Exposure
ICD lead extraction outcomes by manufacturer and lead family
Outcome
Complete procedural success of ICD TLE

Main finding

As reported in the abstract

Complete procedural success, by manufacturer
Complete ICD TLE success, leads (n = 860)
97.2%
Complete ICD TLE success, patients (n = 794)
98.0%
Difference across manufacturers
P = 0.03
Linox family leads, lowest success
89.7%
Linox family vs others
P = 0.02

Observational data: associations, not effects.

Further findings

Predictors of incomplete ICD lead removal (multivariable)

ICD lead age > 10 years and Linox family lead

Predictors of all-cause mortality (multivariable) · within 30 days

Lead extraction for infection, end-stage renal disease, higher NYHA functional class

Safety

Major complications

Major complications
22 patients (2.7%)
Multivariable predictors of major complications: ICD lead age > 15 years and longer lead extraction time
Predictors of all-cause mortality within 30 days: lead extraction for infection, end-stage renal disease, higher NYHA class

Original publication

Extraction outcomes of implantable cardioverter-defibrillator leads vary by manufacturer and model family. Europace. Published 24 November 2023. doi:10.1093/europace/euad345

Summarised from the published abstract; every number is checked against it