Heart failure & cardiomyopathiesResearch article

First-in-Human Implantable Inferior Vena Cava Sensor for Remote Care in Heart Failure

JACC Heart FailPublished

  • First-in-human study

Authors' conclusion

This first-in-human experience demonstrated that the implantation of an IVC sensor was safe and feasible.

Study context

Population
Heart failure (HF) patients with HF hospitalisation within the previous year, elevated natriuretic peptides, on optimal HF treatment49 contributed to primary endpoints
Participants
50 patients
Outcome
Procedural success without device- or procedure-related complications (primary safety); signal acquisition following implantation and at a clinic visit (primary technical)
Follow-up
3 months

Main finding

As reported in the abstract

Successful implantation

50 patients

Further findings

Agreement of sensor-derived inferior vena cava (IVC) area with CT

Mean absolute error
13.53 mm² [3.55%]
R2
0.98

Adherence to daily readings · 6 months

Median adherence
96%

Exploratory clinical outcomes · 6 months

Suggested improvements in NT-proBNP, NYHA class and quality of life and reduced HF events
  • exploratory

Original publication

First-in-Human Implantable Inferior Vena Cava Sensor for Remote Care in Heart Failure. JACC Heart Fail. Published 9 April 2025. doi:10.1016/j.jchf.2025.01.019

Summarised from the published abstract; every number is checked against it