First-in-Human Multicenter Experience of the Newest Generation Supra-Annular Self-Expanding Evolut FX Transcatheter Aortic Valve Replacement (TAVR) System

JACC Cardiovasc IntervPublished

  • Retrospective multicentre cohort study
  • first-in-human
  • limited market release
  • 9 US centres
  • consecutive patients
  • comparison with single-centre PRO+ experience
  • Characteristics retrospectively reviewed

Authors' conclusion

The Evolut FX system demonstrated favorable 30-day outcomes with a significant improvement over PRO+ in achieving commissural alignment, fewer device recaptures, and more symmetrical implantation.

Study context

Population
Patients undergoing transfemoral TAVR for native AS (89.4%) or prosthetic valve degeneration (10.6%)
Participants
226 patients · 9 US centres
Exposure
Evolut FX vs Evolut PRO+
  • FXn = 226
Outcome
Clinical and echocardiographic outcomes (VARC-3 definitions)
Follow-up
30 days

Main finding

As reported in the abstract

Paravalvular leak

Mild paravalvular leak

14.3%

Moderate paravalvular leak

0.9%

Severe paravalvular leak

0

Observational data: associations, not effects.

Further findings

FX vs single-centre Evolut PRO+ experience

P < 0.001

Device recaptures

P = 0.004

Commissural alignment

P < 0.001

  • FX vs single-centre Evolut PRO+ experience: more symmetrical implantation, shallower depth at left coronary cusp

Procedural results

Optimal hat marker orientation during deployment
98.4%
Commissural alignment
96.5%

Original publication

First-in-Human Multicenter Experience of the Newest Generation Supra-Annular Self-Expanding Evolut FX TAVR System. JACC Cardiovasc Interv. Published 10 July 2023. doi:10.1016/j.jcin.2023.05.004

Summarised from the published abstract; every number is checked against it