Arrhythmias & electrophysiologyCohort study

Hemolysis Related to pulsed-field ablation (PFA)

JACC Clin ElectrophysiolPublished

  • prospective analysis
  • consecutive patients

Authors' conclusion

Hemolysis commonly occurs across PFA procedures, irrespective of pulse duration, and is greater with single-shot than focal systems.

Study context

Population
Consecutive patients with atrial fibrillation undergoing first-time PFA67% male, mean age 58.9 ± 11.68
Participants
94 patients
Exposure
hemolysis profiles across different PFA systems
  • Nano PFA 360n = 30
  • Pentasplinen = 34
  • Lattice tipn = 30

Main finding

As reported in the abstract

Hemolysis (>10 mg/dL decrease in Hp between T1 and T2) · 24 hours after the procedure

Hemolysis occurred, all patients

88 (93.6%)

Observational data: associations, not effects.

Further findings

Significant hemolysis (Hp ≤ 25 mg/dL) · 24 hours after the procedure

  • Significant hemolysis (haptoglobin (Hp) ≤ 25 mg/dL): P < 0.001

Significant hemolysis, multivariable analysis

Nano PFA 360 and Pentaspline associated with a similarly high risk of significant hemolysis compared with Lattice tip
P < 0.001
No statistically significant difference between Nano PFA 360 and Pentaspline catheters

Baseline and procedural characteristics

No statistically significant differences among the 3 systems

Original publication

Hemolysis Related to PFA. JACC Clin Electrophysiol. Published 22 July 2026. doi:10.1016/j.jacep.2026.06.022

Summarised from the published abstract; every number is checked against it