Arrhythmias & electrophysiologyResearch article

High-Voltage Nanosecond Pulse Field Ablation Using a Compliant Circular Catheter to Treat Atrial Fibrillation

JACC Clin ElectrophysiolPublished

  • first-in-human study
  • 3 centers

Authors' conclusion

nsPFA demonstrated reasonable safety, good lesion durability, and favorable 1-year clinical effectiveness. (Evaluation of the CellFX® Nano-Pulsed Field Ablation [PFA] 360 Catheter Endocardial Ablation System for the Treatment of Atrial Fibrillation. NCT06696170).

Study context

Population
Patients with symptomatic paroxysmal atrial fibrillationunderwent pulmonary vein isolation (PVI) · 2.5-second applications 36 · 5-second applications 141
Participants
177 patients
Comparison
2.5-second applications vs 5-second applications
  • 2.5-second applicationsn = 36
  • 5-second applicationsn = 141

Main finding

As reported in the abstract

Acute lesion success

All lesions acutely successful

100%

Further findings

PVI durability with 5-second applications · 2 to 3 months
91%

Freedom from atrial arrhythmia · 1 year

1-year estimate for freedom from atrial arrhythmia
89.7% (95% CI: 80.5%-94.6%)

Additional ablation sites

Posterior wall
n = 87
Cavotricuspid isthmus
n = 11
Mitral isthmus
n = 29

Procedural times

Transpired PVI time
12 ± 5 minutes
Total left atrial dwell time for the nanosecond PFA (nsPFA) catheter
19 ± 13 minutes
Total procedure time
61 ± 27 minutes
Fluoroscopy time
9 ± 6 minutes

Safety

Primary serious adverse events

Primary serious adverse events
Three (of 177 [1.7%])
Inflammatory pericardial effusion
Hemolysis with acute kidney injury
Stroke

Post procedure brain MRI findings (35 patients)

Silent cerebral events (DWI+ / ADC-reduced)
11.4%
Silent cerebral lesions (SCE plus FLAIR+)
11.4%
  • among 35 patients with post procedure brain magnetic resonance imaging

Original publication

High-Voltage Nanosecond Pulse Field Ablation Using a Compliant Circular Catheter to Treat Atrial Fibrillation. JACC Clin Electrophysiol. Published 25 April 2026. doi:10.1016/j.jacep.2026.04.012

Summarised from the published abstract; every number is checked against it