Impact of Preoperative Temporary Mechanical Circulatory Support on Durable Left Ventricular Assist Device (LVAD) Outcomes: An Analysis of The Society of Thoracic Surgeons National Intermacs Database
Authors' conclusion
Use of HF-MP and IABP were associated with improved survival and equivalent rate of stroke compared to no tMCS.
Study context
- Population
- Adults receiving a fully magnetically levitated LVAD as Intermacs profiles 1–3, 2020 to 20243,863 (66.7%) on temporary mechanical circulatory support (tMCS)
- Participants
- 5,787 adults
- Exposure
- Preoperative tMCS vs no tMCS before durable LVAD
- Outcome
- Mortality and rates of stroke
- Follow-up
- 1 and 2 years
Methodological detailsregistry-based (STS Intermacs), multivariable proportional hazards model, propensity-score matched
- Registry-based (STS Intermacs)
- Multivariable proportional hazards model
- Propensity-score matched
Main finding
As reported in the abstract
Survival
ECMO
- 1-year survival
- 70.6%
- 2-year survival
- 67.1%
Intra-aortic balloon pump (IABP)
- 1-year survival
- 85.7%
- 2-year survival
- 81.2%
HF-MP
- 1-year survival
- 82.9%
- 2-year survival
- 77.5%
No tMCS
- 1-year survival
- 82.4%
- 2-year survival
- 77%
1-year survival
ECMO70.6%
Intra-aortic balloon pump (IABP)85.7%
HF-MP82.9%
No tMCS82.4%
2-year survival
ECMO67.1%
Intra-aortic balloon pump (IABP)81.2%
HF-MP77.5%
No tMCS77%
Each group vs no tMCSHazard ratio (95% CI)P value
HF-MP
0.71(0.59–0.84)
P = 0.001IABP
0.74(0.63–0.86)
P = 0.001Key notes
- Overall survival 82.9% at 1 year and 77.9% at 2 years
Observational data: associations, not effects.
Further findings
| Outcome | No tMCS | HF-MP | P value |
|---|---|---|---|
| Survival, HF-MP vs no tMCS (PSM cohort) · 1 and 2 yearsMatched cohort | P = 0.009 | ||
| 1-year survival | 80.5% | 85.2% | |
| 2-year survival | 74.3% | 79.9% | |
| Freedom from stroke, HF-MP vs no tMCS (PSM cohort) · 1 yearSimilar · matched cohort | 93.9% | 93.1% | P = 0.33 |
Use of tMCS over the study period
- HF-MP use increased (19.1% to 35.1%)
- P < 0.001
- IABP use decreased (41.2% to 27.3%)
- P < 0.001
Original publication
Impact of Preoperative Temporary Mechanical Circulatory Support on Durable LVAD Outcomes: An Analysis of The Society of Thoracic Surgeons National Intermacs Database. JACC Heart Fail. Published 8 October 2025. doi:10. 1016/ j. jchf. 2025. 102671
Composed by AI from the published abstract and checked against it. How Briefs are made →