Arrhythmias & electrophysiologyResearch article

Improved implantable cardioverter-defibrillator shock efficacy using programmable pulse width

Heart RhythmPublished

  • Intrapatient comparison of shock waveform types using remote monitoring data
  • within-patient comparison

Authors' conclusion

Reprogramming shock waveforms from 65% fixed tilt to fixed pulse width significantly improved intrapatient shock success.

Study context

Population
US patients with single-coil implantable cardioverter-defibrillator (ICD) or CRT-D devices (Ellipse, Assura, Unify, and Fortify, Abbott) who received defibrillation or cardioversion shock therapy when programmed to both 65% fixed tilt and fixed pulse width modes for distinct periodspatients experiencing at least 1 shock-delivered episode with each shock type, 157 [57%] with ICDs, 120 [43%] with CRT-Ds · ICDs 157 · CRT-Ds 120
Participants
277 patients
Comparison
fixed pulse width vs fixed tilt waveforms

Main finding

As reported in the abstract

First-shock success rate

P < 0.001

Further findings

Delivered shock energy, in patients with greater-or-equal first-shock success · comparable energies

P = 0.53

Outcomefixed pulse widthfixed tiltP value
100% first-shock success · more patients with fixed pulse width69.3% of episodes46.2% of patientsP < 0.001

Greater-or-equal first-shock success with fixed pulse width

patients with greater-or-equal first-shock success with fixed pulse width
79.4% of patients (220 of 277)

Original publication

Improved implantable cardioverter-defibrillator shock efficacy using programmable pulse width. Heart Rhythm. Published 27 April 2026. doi:10.1016/j.hrthm.2026.04.035

Summarised from the published abstract; every number is checked against it