Heart failure & cardiomyopathies

In Vivo Fluid Dynamics of the Ventura Interatrial Shunt Device in Patients with Heart Failure

ESC Heart FailPublished Mechanistic study

Authors' conclusion

When implanted in patients with advanced HF, this small interatrial shunt demonstrated predictable and durable patency and performance.

Study context

Population
Open-label patients with heart failure (HF) from the RELIEVE-HF trial
Participants
97 patients
Outcome
In vivo fluid dynamics (effective diameter, trans-shunt gradient, flow) and lumen integrity of the shunt
Follow-up
Implant, 6 and 12 months
Methodological detailsopen-label, serial transesophageal echocardiography (TEE), computational fluid dynamics (CFD) and bench flow tests
  • Open-label
  • Serial TEE
  • CFD and bench flow tests

Main finding

As reported in the abstract

Discharge coefficient (CFD and bench)
0.88 to 0.89
Patency confirmed in all instances except one stenotic shunt at 6 months
1
Deff at 12 months, unchanged from baseline (cm)
0.47 ± 0.01, P = 0.376
Trans-shunt mean pressure gradient at 12 months, unchanged (mmHg)
5.1 ± 3.9, P = 0.316
Shunt flow at 12 months, unchanged (mL/min)
1137 ± 463, P = 0.384
TEE flow vs pressure correlation with fluid dynamics model
R2 ≥ 0.98
Qp/Qs ratio at 12 months
1.22 ± 0.12

Further findings

Implant success

Successfully implanted in all enrolled patients
97
  • Mean age 70 ± 11 years; 97% NYHA III; 51% LVEF ≤40%

Original publication

In Vivo Fluid Dynamics of the Ventura Interatrial Shunt Device in Patients with Heart Failure. ESC Heart Fail. Published 22 May 2024. doi:10.1002/ehf2.14859