Acute cardiac care & resuscitationCohort study

Incidence, implications, and management of sense-B-noise failure in subcutaneous cardioverter-defibrillator patients: insights from a large multicentre registry

EuropacePublished

  • Retrospective multicentre registry study
  • nine centres in Europe and the US

Data from medical records, device telemetry and manufacturer reports

Authors' conclusion

This multicentre analysis' findings shed light on a rare but clinically highly significant adverse event in S-ICD therapy.

Study context

Population
Patients implanted with S-ICD models 1010, A209 and A219, October 2009–July 2023
Participants
1,158 patients
Outcome
Incidence and implications of sense-B-noise events

Main finding

As reported in the abstract

Sense-B-noise failure · median follow-up 46 (IQR 23-64) months
Patients with sense-B-noise failure
6
Proportion of all patients
0.5%
Proportion of all inappropriate shocks (IAS)
5.6%
Diagnosed after an IAS episode
6
Median lead dwell time in affected patients (months)
23 (2-70)

Observational data: associations, not effects.

Further findings

Inappropriate shocks (≥1 IAS)

Patients with ≥1 IAS
107 (9.2%)

Management of sense-B-noise

Reprogramming to secondary vector
3
System removal and S-ICD reimplantation (low secondary-vector amplitude)
2
Secondary vector, then system exchange for T-wave-oversensing IAS
1

Original publication

Incidence, implications, and management of sense-B-noise failure in subcutaneous cardioverter-defibrillator patients: insights from a large multicentre registry. Europace. Published 11 June 2024. doi:10.1093/europace/euae161

Summarised from the published abstract; every number is checked against it