Arrhythmias & electrophysiologyCohort study

Initial clinical experience with the balloon-in-basket pulsed field ablation system: acute results of the VOLT CE mark feasibility study

EuropacePublished

  • Prospective single-arm feasibility study
  • pre-market
  • multicentre
  • feasibility sub-study

Authors' conclusion

The initial clinical use of the Volt PFA System demonstrates acute safety and effectiveness in the treatment of symptomatic, drug refractory AF.

Study context

Population
Patients with paroxysmal or persistent atrial fibrillation (AF)all completed 30-day follow-up
Participants
32 patients
Outcome
Primary serious adverse event rate and acute procedural effectiveness
Follow-up
Within 7 days (serious adverse events)

Main finding

As reported in the abstract

Acute procedural effectiveness · acute

Acute effectiveness, treated pulmonary vein (PVs) (127/128)

99.2%

Acute effectiveness, subjects (31/32)

96.9%

Observational data: associations, not effects.

Further findings

Procedural metrics

Pulsed field ablation (PFA) applications/subject (mean ± SD)
23.8 ± 4.2
Procedure time, min
124.6 ± 28.1
Fluoroscopy time, min
19.8 ± 8.9
LA dwell time, min
53.0 ± 21.0
Transpired ablation time, min
48.0 ± 19.9

Safety

Primary serious adverse events · within 7 days

No primary serious adverse events

Systematic safety assessments

Silent cerebral lesions (3 subjects)
9.4%
No phrenic nerve injury, PV stenosis or oesophageal lesions causally related to the PFA system

Original publication

Initial clinical experience with the balloon-in-basket pulsed field ablation system: acute results of the VOLT CE mark feasibility study. Europace. Published 3 May 2024. doi:10.1093/europace/euae118

Summarised from the published abstract; every number is checked against it