Arrhythmias & electrophysiologyCohort study

Initial real-world data on catheter ablation in patients with persistent atrial fibrillation using the novel lattice-tip focal pulsed-field ablation catheter

EuropacePublished

  • Observational study
  • consecutive patients
  • single-arm
  • General anaesthesia (n = 6) or deep sedation (n = 22)

Authors' conclusion

In terms of real-world data, our data confirmed AF ablation feasibility utilizing the lattice-tip focal PFA/RFA catheter in patients with persistent AF.

Study context

Population
Consecutive patients undergoing catheter ablation for persistent atrial fibrillation (AF)
Participants
28 patients
Outcome
Acute procedural and periprocedural data (feasibility)

Main finding

As reported in the abstract

Additional linear ablations
21 patients (78%)
Successful anterior line
n = 13, 46%
Successful roof line
n = 19, 68%
Median procedural time (IQR: 80-114) min
97
Median fluoroscopic time (IQR: 7.2-9.5) min
8.5

Observational data: associations, not effects.

Further findings

Early recurrent AF within 90-day blanking period

Early recurrent AF documented (incl. 1 during hospital stay)
four patients (15%)
  • 27 patients (96%) interviewed during follow-up

Safety

Major or minor procedural complications

Neither major nor minor procedural complication occurred

Original publication

Initial real-world data on catheter ablation in patients with persistent atrial fibrillation using the novel lattice-tip focal pulsed-field ablation catheter. Europace. Published 25 June 2024. doi:10.1093/europace/euae129

Summarised from the published abstract; every number is checked against it