Arrhythmias & electrophysiology

Integration of PRAETORIAN Score, defibrillation testing and follow-up outcomes for the assessment of conversion failure in patients with subcutaneous implantable cardioverter defibrillator: Results from the iSUSI registry

Heart RhythmPublished registry

Authors' conclusion

PRAETORIAN score, BMI, and impedance are independent predictors of defibrillation failure. Their combined use may improve risk stratification and may help guide clinical decision-making, especially when DFT is omitted.

Study context

Population
Patients with subcutaneous implantable cardioverter defibrillator (S-ICD) with available imaging and complete follow-uppatients analyzed, 748 underwent defibrillation testing (DFT)
Participants
1,063 patients
Exposure
PRAETORIAN score and individual components; multivariable model combining PRAETORIAN score, body mass index (BMI), and impedance vs PRAETORIAN score
Outcome
Composite of:
  • DFT failure
  • Ineffective appropriate shocks during follow-up
Methodological detailsinternational, real-world
  • International
  • Real-world

Main finding

As reported in the abstract

Observational data: associations, not effects.

Further findings

Independent predictors of conversion failure

ComparisonOdds ratio (CI)P value
Anterior generator position at PRAETORIAN score step 22.3 (1.2–4.2)P = 0.006
OutcomeAUCP value
Model performance (AUC)0.78 vs 0.71P = 0.0078
Model performance (AUC) — Overall PRAETORIAN score AUC
0.71
Model performance (AUC) — step 1-3 AUCs
0.68-0.70
Model performance (AUC) — Overall vs individual steps
P > .05 for all comparisons

DFT failure and ineffective shocks

DFT failure among 748 patients who underwent DFT
64 (8.6%)
Ineffective shocks during follow-up
17

Independent predictors of conversion failure

PRAETORIAN score step 1, high body mass index (BMI) and post-shock impedance independently predicted conversion failure

Original publication

Integration of PRAETORIAN Score, defibrillation testing and follow-up outcomes for the assessment of conversion failure in patients with subcutaneous implantable cardioverter defibrillator: Results from the iSUSI registry. Heart Rhythm. Published 26 June 2026. doi:10.1016/j.hrthm.2026.06.024

Summarised from the published abstract; every number is checked against it