Heart failure & cardiomyopathies

Invasive haemodynamic validation of an artificial intelligence echocardiographic tool for detecting heart failure with preserved ejection fraction

ESC Heart FailPublished external validation

Authors' conclusion

This AI model demonstrated high specificity and positive predictive value (PPV) for detecting HFpEF, reliably identifying patients with more advanced haemodynamic abnormalities. Its performance remained robust across variable intervals between transthoracic echocardiography (TTE) and right heart catheterization (RHC), and in patients with indeterminate clinical scores. Due to limited sensitivity, the tool is best utilized to enrich identification of patients with clearly abnormal haemodynamics rather than to exclude HFpEF, particularly in early or borderline cases. While broader use as a screening tool is promising, prospective validation studies are necessary to confirm its utility in general populations.

Study context

Population
Patients with unexplained dyspnoea who underwent transthoracic echocardiography (TTE) and right heart catheterization (RHC)28 with both rest and exercise haemodynamics
Participants
47 patients
Index test or model
FDA-approved artificial intelligence (AI) tool on apical 4-chamber echocardiographic video vs invasive haemodynamic criteria (pulmonary capillary wedge pressure (PCWP) ≥15 mmHg at rest or ≥25 mmHg with exercise)
Target condition
heart failure with preserved ejection fraction (HFpEF)
Methodological detailsretrospective, real-world referral cohort
  • Retrospective
  • Real-world referral cohort

Performance

As reported in the abstract

Sensitivity

30%

Specificity

88%

PPV

86%

NPV

33%

Further findings

OutcomeEstimateP value
Haemodynamics, AI-positive vs AI-negativeAI-positive · significantly higher in AI-positiveResting PCWP 20 vs 15 mmHgP = 0.029
Haemodynamics, AI-positive vs AI-negativeAI-positive · significantly higher in AI-positiveMean PA pressure 29 vs 24 mmHgP = 0.02
Haemodynamics, AI-positive vs AI-negativeAI-positive · significantly higher in AI-positivePVR 2.1 vs 1.3 WUP = 0.002

Heart failure with preserved ejection fraction (HFpEF) by PCWP/CO slope >2 mmHg/L/min (secondary analysis)

Specificity
100%
PPV
100%
Sensitivity
27%
NPV
16%
  • n = 25

Prevalence of haemodynamic HFpEF criteria among patients with exercise RHC (n = 28)

Met haemodynamic HFpEF criteria
71%

Invasively confirmed HFpEF in patients with indeterminate H2FPEF scores (n = 25)

Correctly identified by AI, of those with invasively confirmed HFpEF
80%

Model performance by TTE-RHC interval

Consistent across TTE-RHC intervals <365 and <90 days

Original publication