Arrhythmias & electrophysiology

Left Atrial Appendage Occlusion Using a Novel System: Preclinical Evaluation and First-in-Human Experience

JACC Cardiovasc IntervPublished First-in-human study with preclinical evaluation

Authors' conclusion

This novel LAA occlusion system demonstrated favorable early safety and performance in both preclinical and FIH evaluation.

Study context

Population
Nonvalvular AF patients; preclinical canine modelsplus 17 canines
Participants
10 patients
Outcome
Safety: acute procedural success; performance: mechanical device closure
Follow-up
45 days
Methodological detailssingle-arm, prospective, 3 preclinical canine trials
  • Single-arm
  • Prospective
  • 3 preclinical canine trials

Main finding

As reported in the abstract

Acute procedural success (primary safety endpoint); mechanical device closure (primary performance endpoint)

Acute procedural success (a priori primary safety endpoint)

9 of 9 patients

100%

Mechanical device closure at 45 days (a priori primary performance endpoint)

9 of 9 patients

100%

Observational data: associations, not effects.

Further findings

Successful implantation (first-in-human (FIH))

Successful implantation
9 of 10 patients (90%)

45-day TEE

All implants stable, complete protocol-defined left atrial appendage (LAA) closure and no device-related thrombus (DRT)

Preclinical canine explant analysis · 45 days

Complete endothelialization, no DRT or significant inflammation (n = 17 canines)

Safety

No device- or procedure-related serious adverse events occurred

Original publication

Left Atrial Appendage Occlusion Using a Novel System: Preclinical Evaluation and First-in-Human Experience. JACC Cardiovasc Interv. Published 25 October 2025. doi:10.1016/j.jcin.2025.10.002

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.