Arrhythmias & electrophysiology

Left Bundle Branch Area Defibrillator (LBBA Defibrillator (LBBAD))): A First-in-Human Feasibility Study

JACC Clin ElectrophysiolPublished Prospective feasibility study

Authors' conclusion

This first-in-human evaluation showed the feasibility of LBBAD implantation in a small cohort of patients.

Study context

Population
Patients with an implantable cardioverter-defibrillator (ICD) indicationimplantation achieved in 3
Participants
5 patients
Outcome
Safety and feasibility of implanting an LBBA ICD lead
Methodological detailssingle-centre, first-in-human
  • Single-centre
  • First-in-human

Main finding

As reported in the abstract

Successful LBBAD implantation

LBBAD

60%

3 of 5

Observational data: associations, not effects.

Further findings

Procedural and fluoroscopy duration

Mean procedural duration
170.0 ± 17.3 minutes
Mean fluoroscopy duration
28.8 ± 16.1 minutes

Capture type

Left bundle capture
2 patients (66%)
Left septal capture
1 patient

Paced ECG (left bundle branch area (LBBA) pacing)

Mean QRS duration
121.3 ± 8.3 milliseconds
Mean V6 R-wave peak time
86.1 ± 10.0 milliseconds

Defibrillation testing

Successful defibrillation testing
3
Mean time to adequate shock delivery
8.6 ± 2.6 seconds
  • All 3 patients

Acute LBBA pacing parameters

Pacing threshold
0.80 ± 0.60 V at 0.4 milliseconds
R-wave amplitude
7.0 ± 2.7 mV

Safety

No LBBA lead-related complications

Original publication

Left Bundle Branch Area Defibrillator (LBBAD)): A First-in-Human Feasibility Study. JACC Clin Electrophysiol. Published 22 March 2023. doi:10.1016/j.jacep.2023.01.006

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.