Left bundle branch area stylet-driven lead: Performance, safety, and quality of life at 12 months after implant (the BIO-CONDUCT investigational device exemption study)

Heart RhythmPublished

  • multicenter, prospective, nonrandomized trial

Authors' conclusion

These 12-month results using the Solia S stylet-driven lead demonstrate freedom from LBBA pacing lead-related complications, acceptable lead performance characteristics, and a significant coincident improvement in a patient's physical function.

Study context

Population
Patients with standard pacing indications in whom a Solia S lead was implanted in the left bundle branch area (LBBA)patients who experienced a serious adverse device effect (SADE) event or had at least 335 days of follow-up
Participants
161 patients

Main finding

As reported in the abstract

Pacing threshold · 12 months

Mean threshold at 0.4 ms

0.98 V

Mean threshold at implant

0.84 V

Observational data: associations, not effects.

Further findings

Sensing · 12 months

Mean sensing
12.76 mV
Mean sensing at implant
9.15 mV

Lead impedance · 12 months

Mean impedance
521.8 Ω
Mean impedance at implant
670.5 Ω

Quality of life (QOL) physical function scale · 12 months

Mean change
+11.4 ± 24.0
95% confidence interval
7.4-15.4
t value
4.23
P
P < 0.001

Safety

Serious adverse device effect (SADE) · 12 months

SADE-free rate (95% confidence interval 94.7-99.6)
98.1%
SADE-free rate 0.02 events per subject-year (98.1%; 95% CI, 94.7-99.6)
  • 161 patients who experienced a SADE event or had at least 335 days of follow-up

Original publication

Left bundle branch area stylet-driven lead: Performance, safety, and quality of life at 12 months after implant (the BIO-CONDUCT investigational device exemption study). Heart Rhythm. Published 19 March 2026. doi:10.1016/j.hrthm.2026.03.1889

Summarised from the published abstract; every number is checked against it