Arrhythmias & electrophysiologyCohort study

MAGNETOcardiography Parameters to Predict Sudden Cardiac Death (MAGNETO-SCD) or Ventricular Events From Implantable Cardioverter Defibrillators

JACC Clin ElectrophysiolPublished

  • prospective, multicenter observational study
  • 5 UK hospitals
  • portable, unshielded, room temperature device

Authors' conclusion

In the current study, novel rotation-based MCG parameters were associated with subsequent appropriate ICD therapies, particularly shocks. These exploratory findings support the feasibility of unshielded MCG in routine clinical practice and suggest that Rotation Score/Angular Dynamics may capture arrhythmic substrate and merit further investigation and validation in larger cohorts. (Magnetocardiography [MCG] Parameters in the Prediction of Future ICD Therapy [Magneto-SCD]; NCT04352816; EudraCT2019-002994-78).

Study context

Population
De novo implantable-cardioverter defibrillator (ICD) recipients, September 2019 to November 2021patients recruited, 90 had analyzable magnetocardiography (MCG) data
Participants
104 recipients
Exposure
MCG parameters (Rotation Score and Angular Dynamics) in patients with vs without subsequent appropriate ICD therapy
  • appropriate ICD therapyn = 17
Outcome
Occurrence of any appropriate ICD therapy
Follow-up
Mean 36.0 ± 17.5 months

Main finding

As reported in the abstract

Rotation Score, median (Q1-Q3), appropriate ICD therapy vs not

504.4 [398.5-665.3] vs 331.7 [256.6-472.2]; P = 0.005

Angular Dynamics, median (Q1-Q3), appropriate ICD therapy vs not

3.26 [2.55-3.81] vs 2.50 [2.12-3.13]; P = 0.029

Observational data: associations, not effects.

Secondary outcomes

Outcomehazard (95% CI)
Appropriate shocks — Each 1-SD increase in Rotation Score · higher hazard of appropriate shocks · CIs were wide and event numbers were small2.53 (1.53–4.18)
Appropriate shocks — Each 1-SD increase in Angular Dynamics · higher hazard of appropriate shocks · CIs were wide and event numbers were small2.34 (1.37–4.00)
Appropriate shocks — Rotation Score, with appropriate shocks vs not
539.7 [428.5-674.5] vs 335.1 [256.3-476.6]; P = 0.001
Appropriate shocks — Angular Dynamics, with appropriate shocks vs not
3.348 [2.879-3.830] vs 2.492 [2.117-3.132]; P = 0.005
Appropriate shocks — Receiver-operating characteristic area under the curve for shocks, Rotation Score
0.82
Appropriate shocks — Receiver-operating characteristic area under the curve for shocks, Angular Dynamics
0.77
  • Appropriate shocks: Associations persisted after limited adjustment for age and left ventricular ejection fraction; requires independent validation

Original publication

MAGNETOcardiography Parameters to Predict Sudden Cardiac Death (MAGNETO-SCD) or Ventricular Events From Implantable Cardioverter Defibrillators. JACC Clin Electrophysiol. Published 2 July 2026. doi:10.1016/j.jacep.2026.05.020

Summarised from the published abstract; every number is checked against it