Measuring Congestion with a Non-Invasive Monitoring Device in Heart Failure and Haemodialysis: CONGEST-HF
Authors' conclusion
A non-invasive device was able to detect changes in congestion in patients with HF receiving decongestion therapy and patients having fluid removed at haemodialysis.
Study context
- Population
- Patients with heart failure (HF) or on haemodialysis assessed with a wearable cardiopulmonary sensor (worn <5 min)
- Participants
- 66 patients3 cohorts
- Exposure
- HF vs Haemodialysis vs right heart catheterization (RHC)
- HFn = 25
- Haemodialysisn = 21
- RHCn = 20
- Outcome
- Spearman correlation per cohort: (1) change in weight vs change in thoracic impedance (TI) in HF; (2) lung ultrasound B-lines and dialysis volume removed vs device TI; (3) PCWP vs device S3 in RHC
Methodological details3 cohorts
- 3 cohorts
Main finding
As reported in the abstract
- HF: change in weight vs change in device thoracic impedance (TI), rsp
- −0.64, p = 0.002
- HF: change in weight vs change in device S3, rsp
- −0.53, p = 0.014
- Haemodialysis: B-lines vs TI before dialysis, rsp
- −0.71, p < 0.001
- Haemodialysis: B-lines vs TI after dialysis, rsp
- −0.77, p < 0.001
- Haemodialysis: volume removed vs change in device TI, rsp
- 0.49, p = 0.024
- RHC: Pulmonary capillary wedge pressure (PCWP) (one time point) vs device S3, not significant, rsp
- 0.230, p = 0.204
Observational data: associations, not effects.
Safety
- No device-related adverse events
Original publication
Measuring Congestion with a Non-Invasive Monitoring Device in Heart Failure and Haemodialysis: CONGEST-HF. Eur J Heart Fail. Published 13 May 2024. doi:10. 1002/ ejhf. 3290