Valvular & structural heart diseaseSecondary analysis

Mechanisms, predictors, and evolution of severe peri-device leaks with two different left atrial appendage occluders

EuropacePublished

  • Secondary analysis of the Amulet IDE randomised trial
  • independent core lab
  • propensity-score matched controls

Authors' conclusion

The dual-occlusive Amplatzer Amulet LAA occluder provided improved LAA closure compared with the Watchman 2.5 device. Predictors and temporal observations of severe PDLs were identified in the Amulet IDE trial.

Study context

Population
Patients randomised in the Amulet IDE trial undergoing left atrial appendage occlusion (LAAO)randomised
Participants
1,878 patients
Comparison
Amulet vs Watchman 2.5 LAAO devices
Primary endpoint
Severe peri-device leak (>5 mm): incidence, mechanisms, predictors and evolution
Follow-up
Through 18 months

Primary result

As reported in the abstract

Severe peri-device leak (>5 mm)

45 days

P < 0.001

12 months

P < 0.001

Further findings

Evolution of Watchman severe leaks · through 18 months

Most Watchman severe leaks did not resolve over time

Original publication

Mechanisms, predictors, and evolution of severe peri-device leaks with two different left atrial appendage occluders. Europace. Published 16 August 2023. doi:10.1093/europace/euad237

Summarised from the published abstract; every number is checked against it